Installation Instructions for the TekRxTest_Application Software Product: TekRxTest_Application Version: 6.1.0.x Date: Jan-2021 ================================================================================================================ Installation Instruction 1 Upgrade • It is mandatory to uninstall existing beta application (If any) before installing a new version • Uninstall “TekRxTest_Application” and “TekRxService” from Control Panel • If TekRxservice ("SigTestService" from older versions) launches through Start-Up service, then remove from the path C:\ProgramData\Microsoft\Windows\Start Menu\Programs\Startup • Check if the folders from C:\Program Files\Tektronix\BERTScope are removed after uninstallation • Rename Python2.7 folder in C: drive if the folder is not deleted after uninstallation • Launch the installer wrapper by clicking on TekRxTest_Applica_V6.1.0.x executable • The wrapper contains TekRxService V2.8.0.X , SigTest 4.0.x, SigtestPhoenix 5.0.x and TekRxTest Application V6.1.0.X • Select all the components and click "Next" • Proceed as instructed 2 Fresh Installation • The installation wrapper contains TekRxService V2.8.0.X , SigTest 4.0.x, SigtestPhoenix 5.0.x and TekRxTest Application V6.1.0.X • Select all the components and click "Next" • Proceed as instructed 3 Desktop icons • Once installation is complete, "TekRxTest" shortcut is created • "TekRxService" batch file shortcut is created 4 Handling connectivity challenges with TekRxService In case there is difficulty in connecting with the TekRxService from the TekRxTest SW, the following checks can be carried out: • Make sure that while launching the TekRxService, the appropriate firewall permissions are granted • If TekRxTest SW is not able to establish the connection with TekRxService after launching it, then port number configuration needs to be performed in Windows firewall settings as follows: Go to Control Panel > System and Security > Windows Firewall Click on advanced Setting > Inbound Rules > New Rules In New inbound rules wizard select Port and click Next > Select specific Ports: 4006 > Click Next (Allow the connection) > Click Next (Domain, Private and Public should be checked) > Name: TekRxService (Save with any name) > Click Finish. • After configuring Firewall setting, launch TekRxService from desktop • Ensure connection is established and the port number is displayed in the TekRxService window 5 Perform SPC of the scope channel before start of Calibration/Tests if required(SPC warning is displayed) • Remove all active signals from the Oscilloscope Channels • Use the scope SPC software to perform the calibration (Utilities > Instrument Calibration > Run SPC) • Check to make certain that SPC warning is no longer displayed 6 Storage of reports • The reports are stored in the following path C:\ProgramData\Tektronix\BERTScope\RxTestApp\PCIE 5.0 • In case the reports are required to be viewed on a separate device as in another RT Scope or BERT Scope or a laptop, copy all the folders and save it in the same location as above but on the viewing device • Launch the TekRxTest SW and notice that all the reports get listed in TP1 and TP2 wizard • Click the “Report” button after selecting the desired report for viewing ===================================================================================================================== If you are unable to download or install this software, please contact the Tektronix Technical Support Center: Customers calling within the U.S. call 1-800-833-9200 press 3. Customers calling from outside the U.S. call 503-627-2400.